对生物组织氧浓度的连续监测对于自适应治疗和长期植入式诊断至关重要。本文提出一种异质集成的光致发光氧传感系统,将CMOS单光子雪崩二极管(SPAD)与可直接键合在芯片上的、脉冲与电流可调的微米级LED(μLED)相结合。采用光学活性聚合物的一步封装工艺同时实现了传感层与保护层的一体化。该系统在高达6.6 ksps的采样率下实现0.03至0.5 % O2的浓度分辨率,单次测量能耗为73.4 nJ,相较现有最先进系统在精度上提升3倍、采样率提升20倍,且功耗仅为其三分之一。
Continuous monitoring of oxygen concentration in biological tissue is critical for adaptive therapies and long-term implant diagnostics. This work presents a heterogeneously integrated photoluminescent oxygen sensing system combining CMOS single-photon avalanche diodes (SPADS) with pulse- and current-controllable micrometer-sized LEDs (μLEDs) die-bonded onto the chip. A single-step encapsulation using an optically active polymer integrates both the sensing and protecting layers. The system achieves a concentration resolution of 0.03 to 0.5 % O2 with an energy consumption of 73.4 nJ per measurement at up to 6.6 ksps, outperforming state-of-the-art systems by 3x in accuracy, 20x in sampling rate, and requiring only one-third of the power.
| 指标 | 数值 | 说明 |
|---|---|---|
| 氧浓度分辨率 | 0.03–0.5 % O2 | 覆盖生理(5–8 %)与乏氧(<2 %)区间的可分辨浓度变化 |
| 单次测量能耗 | 73.4 nJ | 完成一次激励+磷光寿命测量所消耗的能量 |
| 最高采样率 | 6.6 ksps | 连续测量模式下可实现的最大采样速率 |
| ASIC芯片面积 | 2 × 2.5 mm² | 180 nm CMOS 工艺实现的含SPAD与LED驱动芯片面积 |
| 微型化系统尺寸 | 4.5 × 4 × 2 mm³ | 去除测试用连接器与无源件、仅保留柔性板后的封装系统体积 |
| LED驱动PWM频率 | 2 MHz | 通过直接短接驱动电流镜实现的最大脉冲调制频率 |
氧浓度分辨率 0.03–0.5 % O2;单次测量能耗 73.4 nJ(最高 6.6 ksps)
异构集成、模拟前端集成、近阈值低功耗
植入式医疗监测、肿瘤乏氧与放疗适配、组织氧浓度实时监测
打开交互版 · AI 解读 / 同类对比 / 完整指标在香农学术库中查看(需会员)