本文首次提出在0.18μm BCD平台上协同设计的64通道宽带PMUT阵列与多电平CMOS发射机(TxCMOS),在仅±5V驱动电压下于5MHz实现创纪录的70%带宽。我们创新性地将多PMUT通道架构与新型多电平脉冲发生器集成,为可穿戴应用带来高收发效率,其中发射效率达8.4 kPa/V/mm²,自接收灵敏度为1.2 mV/V。与CMOS低噪声放大器(LNA)集成后,接收信号被提升至60mVpp(6倍电压增益)。尤为关键的是,我们首次通过TxCMOS-PMUT实现了人体深部动脉成像,采集到肱动脉(直径5 mm,深度15 mm)的M型超声图像。该系统在上下动脉壁同步跟踪中达到19 µm分辨率,从而可计算与心动周期相关的血压。这一突破使宽带TxCMOS-PMUT成为可穿戴、无创心血管监测与高分辨率深部组织成像的变革性技术。
This paper presents the first co-designed 64-channel broadband PMUT array and multi-level CMOS transmitter (TxCMOS) in 0.18μm BCD platform, achieving a record 70% bandwidth at 5MHz with only ±5V drive voltage. Our innovative integration of multi-PMUT channel architecture and novel multi-level pulser enables high transceiving efficiency for wearables, including 8.4 kPa/V/mm² for transmission and 1.2 mV/V for self-reception sensitivity. Integration with a CMOS low-noise amplifier (LNA) boosts received signals to 60mVpp signals (6x voltage gain). Critically, we demonstrate—for the first time—human deep artery imaging via TxCMOS-PMUT, capturing M-mode ultrasound imaging of the brachial artery (5 mm diameter at 15 mm depth). The system achieves 19 µm resolution in simultaneous upper and lower arterial wall tracking, enabling cardiac-cycle-correlated blood pressure calculation. This breakthrough establishes the broadband TxCMOS-PMUT as a transformative technology for wearable, non-invasive cardiovascular monitoring and high-resolution deep-tissue imaging.
| 指标 | 数值 | 说明 |
|---|---|---|
| 系统带宽 | 70% @ 5 MHz(±5V驱动) | TxCMOS-PMUT系统时域脉冲回波测试达到的-6dB相对带宽 |
| 发射效率 | 8.4 kPa/V/mm² | 在3.27 mm深度由针式水听器测得的单位电压单位面积声压 |
| 自接收灵敏度 | 1.2 mV/V @ 6.5 mm | 以水面为反射体、相邻PMUT通道作接收的脉冲回波灵敏度 |
| 接收增益 | 6倍(60 mVpp) | 集成CMOS LNA后回波信号由11 mVpp提升至60 mVpp |
| 成像分辨率 | 19 µm | 15 mm深度肱动脉上下壁同步跟踪的最小位移分辨率 |
| 单通道PMUT带宽 | 81.7% -6dB | 2周期10Vpp、5MHz正弦脉冲激励下水中的脉冲回波通道带宽 |
5MHz下70%带宽(±5V驱动);发射效率8.4 kPa/V/mm²
低温工艺、图案化与光刻
可穿戴医疗超声、心血管无创监测、深部组织高分辨成像
打开交互版 · AI 解读 / 同类对比 / 完整指标在香农学术库中查看(需会员)