植入式刺激器对无线供电的需求日益增长,但现有电感、超声、磁电、光学和中场射频链路无法在真正可由针头递送的几何尺寸下同时实现高效率、对准鲁棒性和深部神经靶点。组织阻抗随频率和位置强烈变化、姿态引起的耦合漂移以及不断变化的刺激负载,进一步使传统能量采集器难以维持稳定供电。我们提出高动态范围可注射差分组织耦合供电(HDR-DTCP),这是一种高频体耦合技术:皮肤表面的差分电极驱动交流信号通过组织,线状植入体差分拾取该信号并进行片上高动态范围整流。由于耦合分布于植入体长度而非其横截面,HDR-DTCP 具有随长度可扩展的功率传输效率,从而实现有利于微创注射的线状形态因子,这是线圈、叠层或换能器方案目前无法达到的。所提出的 65nm 原型集成了可重构 4 级能量采集器,在 –16.3 dBm 下实现 63.5% 的峰值功率转换效率(PCE),并在 9 dB 输入范围内保持大于 50% 的 PCE;同时集成了数字可调双相刺激器,可驱动高达 RL > 1 kΩ 和 CL < 20 pF 的负载,朝着完全可注射神经刺激器迈进。
Implantable stimulators increasingly demand wireless power, yet prevailing inductive, ultrasonic, magnetoelectric, optical, and mid-field RF links cannot simultaneously achieve high efficiency, alignment robustness, and deep nerve target in a truly needle-deliverable geometry. Strongly frequency and site-dependent tissue impedances, posture-driven coupling shifts, and changing stimulation loads further prevent conventional energy harvesters from maintaining a stable supply. We present High-Dynamic-Range Injectable Differential Tissue-Coupled Powering (HDR-DTCP), a high-frequency body-coupled technique where skin-surface differential electrodes drive AC through tissue and a thread-like implant differentially picks it up for on-chip HDR rectification. Because coupling is distributed over the implant length rather than its cross-section, HDR-DTCP yields length-scalable power transfer efficiency, enabling a thread-like form factor conducive to minimally invasive injection methods that are currently unattainable with coils or laminate or transducer-based approaches. The proposed 65 nm prototype integrates a reconfigurable 4-stage harvester that achieves 63.5% peak power conversion efficiency (PCE) at –16.3 dBm and sustains > 50% PCE over a 9 dB input range, together with a digitally tunable biphasic stimulator capable of driving loads up to RL > 1 kΩ and CL < 20 pF, pointing toward fully injectable neurostimulators.
| 指标 | 数值 | 说明 |
|---|---|---|
| 峰值功率转换效率(PCE) | 63.5% @ –16.3 dBm | 可重构4级能量采集器在–16.3 dBm输入下的峰值整流效率 |
| PCE动态范围 | 9 dB 内 > 50% | 输入功率变化9 dB范围内PCE仍保持在50%以上 |
| 刺激器驱动负载能力 | RL > 1 kΩ, CL < 20 pF | 数字可调双相刺激器可驱动的阻性与容性负载上限 |
| 工艺节点 | 65 nm | 集成能量采集器与刺激器的原型芯片工艺 |
峰值PCE 63.5% @ –16.3 dBm;9 dB输入范围内PCE > 50%
电源完整性、数字辅助模拟、模拟前端集成
植入式神经刺激、微创可注射神经调制、无电池生物电子植入体
打开交互版 · AI 解读 / 同类对比 / 完整指标在香农学术库中查看(需会员)