本文提出了一种用于植入式系统的同时16通道绝热电荷跟踪刺激器(SACT)。时分复用刺激方式使单个电流产生模块即可实现并行工作。片上电容保证每一通道精确的电荷平衡,而电感则在每个周期对能量进行绝热回收。基于时间的传感器用于预测相位切换时刻。在500 Ω–10 kΩ负载与60–250 nC刺激电荷范围内,实现了87.4%的刺激器效率、小于3.7%的效率变化以及小于0.54%的电荷失配。并成功演示了体内多通道刺激。
This paper presents a simultaneous 16-channel adiabatic charge-tracking stimulator (SACT) for implantable systems. Time-multiplexed stimulation enables parallel operation using a single current-generation block. An on-chip capacitor ensures accurate per-channel charge balance, while an inductor adiabatically recovers energy each cycle. A time-based sensor predicts phase transitions. A stimulator efficiency of 87.4% with <3.7% variation and <0.54% charge mismatch across 500 Ω–10 kΩ and 60–250 nC stimulation charge was achieved. In-vivo multichannel stimulation was successfully demonstrated.
| 指标 | 数值 | 说明 |
|---|---|---|
| 刺激器效率 | 87.4% | 在500 Ω–10 kΩ与60–250 nC范围内测得的能量转换效率 |
| 效率变化 | <3.7% | 跨电极-组织界面阻抗与刺激电荷范围的效率波动 |
| 电荷失配 | <0.54% | 各通道双相刺激周期的电荷平衡精度 |
| 通道数 | 16通道/32电极 | 由单个电流产生模块经时分复用驱动 |
| 刺激电荷范围 | 60–250 nC | 测试覆盖的每周期刺激电荷区间 |
| 负载范围 | 500 Ω–10 kΩ | 电极-组织界面阻抗测试范围 |
刺激器效率 87.4%;电荷失配 <0.54%
时分复用、数字辅助模拟、自校准与修调
植入式神经刺激器、多通道神经假体、体内多通道电刺激验证
打开交互版 · AI 解读 / 同类对比 / 完整指标在香农学术库中查看(需会员)