本文提出了首个用于多模态神经信号监测的无线植入式SoC。该系统在功耗受限的无线供电平台上集成了快速扫描循环伏安法(FSCV)与基于有机电化学晶体管(OECT)的局部场电位(LFP)传感。该SoC包含:一个由PID控制的背景扣除环路,用于针对FSCV快速变化的背景电流实现高速动态范围扩展;一个无量化噪声的任意波形发生器与一个可重构恒电位仪,从而在不使用片外元件的情况下支持双模态传感器激励;以及一个高能效的基于整流器的动态电压供给(DVS),用于缓解低噪声但高功耗恒电位仪所带来的功耗瓶颈。该SoC采用65nm CMOS工艺制造,在10Hz FSCV扫描时功耗为8.2μW,同时可提供从-0.45V到1.8V的宽传感器偏置电压范围。其动态范围扩展环路的响应速度比现有技术至少快3倍,同时保持FSCV的105.3dB与OECT的138.8dB的动态范围。体外验证证实,在400V/s扫描速率下对多巴胺的灵敏度为208nA/nMol。
This work presents the first wireless implantable SoC for multimodal neural signal monitoring. The system integrates fast-scan cyclic voltammetry (FSCV) and organic electrochemical transistor (OECT)-based local field potential (LFP) sensing within a power-constrained, wirelessly powered platform. The SoC comprises a PID-controlled background subtraction loop to enable high-speed DR expansion for FSCV's fast-varying background currents, a quantization-noise-free arbitrary waveform generator and a reconfigurable potentiostat to support dual-mode sensor excitation without off-chip components, and an energy-efficient rectifier-based dynamic voltage supply (DVS) to mitigate the power bottleneck of the low-noise but power-hungry potentiostat. Fabricated in a 65nm CMOS process, the SoC consumes 8.2μW during 10Hz FSCV scans, while enabling a wide sensor bias voltage range from -0.45V to 1.8V. It achieves at least 3× faster DR expansion loop response than prior techniques while preserving dynamic ranges of 105.3dB for FSCV and 138.8dB for OECT. In vitro validation confirms 208nA/nMol dopamine sensitivity at 400V/s scan rate.
| 指标 | 数值 | 说明 |
|---|---|---|
| 功耗 | 8.2μW | 10Hz FSCV扫描时整芯片功耗 |
| 动态范围(FSCV) | 105.3dB | 快速扫描循环伏安法读出通道的动态范围 |
| 动态范围(OECT) | 138.8dB | OECT电生理模式读出的动态范围 |
| 多巴胺灵敏度 | 208nA/nMol | 400V/s扫描速率下体外实测灵敏度 |
| 传感器偏置电压范围 | -0.45V ~ 1.8V | 可重构恒电位仪支持的宽偏置输出范围 |
| DR扩展环路响应速度 | ≥3× | 相比已有阈值环、TNS-BS等技术的响应速度提升 |
8.2μW功耗;FSCV动态范围105.3dB、OECT动态范围138.8dB
亚阈值设计、背景校准、数字辅助模拟、模拟前端集成
植入式神经疾病长期监测、神经递质(多巴胺)检测、慢性神经系统疾病研究
打开交互版 · AI 解读 / 同类对比 / 完整指标在香农学术库中查看(需会员)