本文提出了一款全集成、鲁棒且极低功耗的、基于卷积神经网络(CNN)的心音图(PCG)监测系统级芯片(SoC)。它包含三项关键技术:(1)动态分帧机制结合片上训练,在存在个体内与个体间变异的情况下提升诊断准确率与能效;(2)可重构计算阵列(RCA)同时支持滤波、动态分帧、推理与训练,从而提高硬件利用率;(3)针对模拟前端(AFE)与CNN计算的全面时钟/功率门控方案进一步提升能效。所提出的28nm PCG SoC在Yassen数据集上实现了98.9%的分类准确率。此外,它在多种生理状态下取得97.1%的平均准确率与0.72%的标准差,同时每次分类仅消耗0.36 μJ。
This paper presents a fully integrated, robust, and extremely low-power phonocardiogram (PCG) monitoring System-on-Chip (SoC) that utilizes convolutional neural networks (CNNs). It incorporates three key techniques: (1) a dynamic framing mechanism combined with on-chip training enhances diagnostic accuracy and energy efficiency despite intra- and inter-individual variability; (2) a reconfigurable computing array (RCA) simultaneously supports filtering, dynamic framing, inference, and training, thereby improving hardware utilization; (3) a comprehensive clock/power-gating scheme for both analog front-end (AFE) and CNN computations further improves energy efficiency. The proposed 28-nm PCG SoC achieves 98.9% classification accuracy on the Yassen dataset. Additionally, it attains an average accuracy of 97.1% with a standard deviation of 0.72% across diverse physiological states, while consuming 0.36 μJ per classification.
| 指标 | 数值 | 说明 |
|---|---|---|
| 分类准确率 | 98.9% | 在Yassen数据集上的五分类诊断准确率 |
| 单次分类能耗 | 0.36 µJ/classification | 整芯片每次分类的平均能量消耗 |
| 跨生理状态平均准确率 | 97.1% | 多种生理状态下平均准确率,标准差0.72% |
| 动态分帧功耗降低 | 71% | 相比固定DTW基线在24小时心率变异性测试中的计算功耗下降 |
| 动态分帧延迟降低 | 82% | 相比固定DTW基线的处理延迟下降,且精度保持不变 |
98.9% 分类准确率;0.36 µJ/classification
机器学习、深度学习、模拟前端集成
可穿戴心脏健康监测、瓣膜性心脏病早期筛查、长期心血管状态监护
打开交互版 · AI 解读 / 同类对比 / 完整指标在香农学术库中查看(需会员)